Before you start
- Make a master mix for all reactions plus 10% extra.
- Controls: known wild-type, heterozygous and homozygous DNA, and a no-template control.
Reagents and materials
Totals are for 24 reactions (+10% extra). Per-reaction (20 µL) amounts are in the middle column.
| Reagent | Per reaction (20 µL) | Total |
|---|---|---|
| 2× PCR master mix (Taq, dNTPs, MgCl₂, buffer) | 10 µL | 10 µL |
| Forward primer, 10 µM (0.5 µM final) | 1 µL | 1 µL |
| Reverse primer, 10 µM (0.5 µM final) | 1 µL | 1 µL |
| Nuclease-free water | 6 µL | 6 µL |
| Genomic DNA templateadded separately to each tube | 2 µL | 2 µL |
Equipment
- Thermocycler
- PCR tubes/plate
- Filter tips
- Gel electrophoresis set-up
Procedure
Tick steps as you go (saved in this browser). Steps with a clock have a built-in timer.
Make the master mix
Combine master mix, primers and water on ice; vortex briefly and spin down.
Dispense
Add 18 µL master mix to each tube, then 2 µL template; mix and spin.
Cycle
Typical program: 94–95 °C 3 min; 35 cycles of 94 °C 30 s, 55–62 °C 30 s (primer-specific), 72 °C 30–60 s per kb; final 72 °C 5 min; hold at 4 °C.
Run a gel
Load 10 µL on a 1.5–2% agarose gel with a ladder and run until bands separate.
Call genotypes
Compare band sizes with your controls and the expected sizes for each allele.
Critical points
- Use the annealing temperature validated for your primer set (often from the strain’s published protocol).
- Set up reactions in a separate area from gel/PCR product handling.
Troubleshooting
| Problem | What to check or change |
|---|---|
| No bands | Check template amount, primer integrity, MgCl₂, annealing temperature (try a gradient). |
| Extra bands | Raise the annealing temperature, lower cycle number or template. |
| Band in no-template control | Replace water and primers; clean pipettes; use filter tips. |
Safety
Standard laboratory care.
Further reading
- Jackson Laboratory strain genotyping protocols (strain-specific primers and conditions).
This is a general starting-point protocol written for SolvesEverything. Follow your institution’s approved protocols, safety rules and manufacturers’ instructions, and optimise for your system. Spotted an error? Tell us.