Molecular biology

DNA and RNA quantification (NanoDrop and Qubit)

Measure nucleic-acid concentration and purity by UV absorbance, and use fluorescent dyes when accuracy matters.

Total time10–20 min
Hands-on10 min
LevelBasic
Steps4
0 of 4 steps done

Before you start

  • Blank with the same buffer your sample is in.

Reagents and materials

Totals are for 12 samples (+10% extra). Per-sample amounts are in the middle column.

ReagentPer sampleTotal
Sample (NanoDrop)1.5 µL1.5 µL
Qubit working solution (dye 1:200 in buffer)199 µL199 µL
Sample (Qubit)1–20 µL; adjust buffer to 200 µL1 µL1 µL

Buffers and recipes

Absorbance conversions (A260 = 1.0, 1 cm path) per reference

Double-stranded DNA50 µg/mL
RNA40 µg/mL
Single-stranded DNA33 µg/mL
Pure DNA A260/280≈ 1.8
Pure RNA A260/280≈ 2.0
A260/230 (both)≈ 2.0–2.2; low = salt, guanidine, phenol

Equipment

  • NanoDrop or other UV spectrophotometer
  • Qubit fluorometer and assay tubes

Procedure

Tick steps as you go (saved in this browser). Steps with a clock have a built-in timer.

  1. Clean pedestals

    Clean the upper and lower pedestals with water and a lint-free wipe.

  2. Blank

    Load 1–2 µL of the sample buffer and blank the instrument.

  3. Measure

    Load 1–2 µL of each sample; record concentration and A260/280 and A260/230 ratios. Measure twice for low concentrations.

  4. Qubit (if accuracy matters)

    Mix dye working solution with sample to 200 µL, incubate 2 min at room temperature and read against the kit standards.

Critical points

  • UV absorbance cannot tell DNA from RNA or free nucleotides; Qubit dyes are specific.
  • Below about 10 ng/µL, NanoDrop readings are unreliable.

Troubleshooting

ProblemWhat to check or change
Odd ratiosRe-blank, clean the pedestal, check for contaminants; re-purify if needed.
Readings differ between methodsTrust the dye-based value for library prep and sequencing.

Safety

Standard laboratory care.

Further reading

  • Instrument manuals (NanoDrop, Qubit).

This is a general starting-point protocol written for SolvesEverything. Follow your institution’s approved protocols, safety rules and manufacturers’ instructions, and optimise for your system. Spotted an error? Tell us.