BCA protein assay (microplate)

Measure protein concentration against a BSA standard curve so you can load equal protein on gels.

Total time45 min
Hands-on15 min
LevelBasic
Steps6
0 of 6 steps done

Before you start

  • Dilute samples so they fall within the standard range (RIPA lysates are often diluted 1:5–1:10).

Reagents and materials

Totals are for 40 wells (+10% extra). Per-well amounts are in the middle column.

ReagentPer wellTotal
Standard or sample25 µL25 µL
BCA working reagent (50 parts A : 1 part B)200 µL200 µL

Buffers and recipes

BSA standard series (µg/mL) per plate

A2000
B1500
C1000
D750
E500
F250
G125
H25
Blank0 (lysis buffer only)

Prepare standards in the same buffer as the samples.

Equipment

  • 96-well plate
  • 37 °C incubator
  • Plate reader at 562 nm

Procedure

Tick steps as you go (saved in this browser). Steps with a clock have a built-in timer.

  1. Prepare standards

    Make the BSA series in the same buffer as your samples.

  2. Load

    Pipette 25 µL standards and samples in duplicate or triplicate.

  3. Add working reagent

    Add 200 µL working reagent and mix on a plate shaker for 30 s.

  4. Incubate

    Cover and incubate at 37 °C for 30 min.

  5. Read

    Cool to room temperature and read at 562 nm.

  6. Calculate

    Subtract the blank, fit the standard curve (linear or quadratic) and calculate sample concentrations × dilution factor.

Critical points

  • Reducing agents (DTT, β-mercaptoethanol) and chelators interfere with BCA; use a compatible kit or Bradford.
  • Colour keeps developing slowly — read all wells at the same time.

Troubleshooting

ProblemWhat to check or change
Samples above rangeDilute further and repeat.
Purple blankInterfering substance in the buffer; dilute or use a compatible assay.

Safety

Reagent A is alkaline; reagent B contains copper sulfate.

Further reading

  • Smith PK et al. Measurement of protein using bicinchoninic acid. Anal Biochem (1985).